Categories

QE65000 Spectrometer

QE65000 Spectrometer

QE65000 Spectrometer

QE65000

Description

The Ocean Optics QE65000 Spectrometer includes the linear CCD-array optical bench, plus all the circuits necessary for spectrometer operation. The result is a compact, flexible system, with no moving parts, that's easily integrated as an OEM component

 

 The QE65000 spectrometer is a unique combination of technologies providing users with both an unusually high spectral response and high optical resolution in a small footprint. The electronics have been designed for considerable flexibility in connecting to various modules as well as external interfaces. The QE65000 interfaces to PCs, PLCs and other embedded controllers through USB 2.0 or RS-232 communications. The information included in this data sheet provides detailed instructions on the connection and operation of the QE65000.

The detector used in the QE65000 spectrometer is a scientific-grade, back-thinned, TE Cooled (TEC), 1044x64 element CCD array from Hamamatsu (product number S7031-1006). For complete details on this detector, visit www.Hamamatsu.com.

The QE65000 operates from power provided through the USB (without TEC operation), or from a separate + 5VDC power supply. The QE65000 is a microcontroller-controlled spectrometer, thus all operating parameters are implemented through software interfacing to the unit.

 

 

Features

 Hamamatsu S7031-1006 Detector:

  • Peak QE: 90%
  • Back-thinned for good UV sensitivity
  • MPP operation for low noise operation
  • TE Cooled

Spectrometer Design:

  • Symmetrical Crossed Czerny Turner
  • 101mm focal length
  • 14 gratings (H1 – H14)
  • 6 slit widths

Electrical Performance:

  • 16 bit, 500KHz A/D Converter
  • Integration times from 8ms to ~10 minutes

Embedded microcontroller allows programmatic control of all operating parameters and standalone operation:

  • USB 2.0 480Mbps (high-speed) and 12Mbps (full speed)
  • RS232 115Kbaud
  • Multiple Communication Standards for digital accessories (SPI, I2C)

TE Cooler

  • Software-controlled set-point
  • Temperature stability: <0.1ºC
  • Maximum cooling 40ºC below ambient

Onboard Pulse Generator:

  • 2 programmable strobe signals for triggering other devices
  • Software control of nearly all pulse parameters

Onboard GPIO:

  • 10 user-programmable digital I/O

EEPROM storage for:

  • Wavelength Calibration Coefficients
  • Linearity Correction Coefficients
  • Absolute Irradiance Calibration (optional)

Plug-n-Play Interface for PC applications

30-pin connector for interfacing to external products


 

 

Specifications:

Specifications

Criteria

Physical Specifications:

Physical Dimensions (LxWxH)

Spectrometer Weight

Power Supply Weight

182 mm (7.17 in.) x 110 mm (4.33 in.) x 47 mm (1.85 in.)

1.15 kg (2.6 lbs.)

0.45 kg (1 lb.)

Power:

Power requirement

Supply voltage

Power-up time

GPIO Output Voltage

500 mA at +5 VDC (no TE Cooler) 3.5A at +5V (with TE Cooler)

4.5 – 5.5 V

~2s depending on code size

3.3 V

Absolute Maximum Ratings:

VCC

Voltage on GPIO pin

+ 5.5 VDC

3.6 V

Spectrometer:

Design

Focal length (input)

Input Fiber Connector

Gratings

Entrance Slit

Detector

Pixels

Pixel size

Active area

Spectral range

Quantum efficiency

Well Depth

Dark Current

Filters

Symmetric crossed Czerny-Turner

F/4 101mm

SMA 905 to single-strand optical fiber (0.22 NA)

14 different gratings (H1 – H14); HC1 grating

5, 10, 25, 50, 100, or 200 μm slits. (Slits are optional. In the absence of a slit, the fiber acts as the entrance slit.)

Hamamatsu S7031-1006

1024 x 58 (total pixels 1044 x 64)

24.6μm square

24.576 x 1.392mm

200 – 1100nm with window (see etalon section). VUV option available (150nm). VUV option includes purge port & window removal (no TE Cooling)

90% peak (see figure following this table)

Individual pixel: 300K electrons; Column summing well: ~1.5M electrons

4000 e/pixel/sec (typ) @ 25° C; 200 e/pixel/sec (typ) @ 0° C

6 OF Series available

Spectroscopic:

Integration Time

Dynamic Range

Signal-to-Noise

Dark Noise (single dark spectrum)

Readout Noise

Linearity

8ms – 15 minutes

40000:1 single integration period, 7.5 x 109 over integration range

1000:1 single acquisition

3 counts RMS

1.5 counts RMS (40 electrons)

~1% max. nonlinearity (uncorrected); <0.5% nonlinearity (corrected)

Specifications

Criteria

Environmental Conditions:

Temperature TEC Control Temperature

Humidity

0 to 50°C 45 to 30°C below ambient temperature

0% – 90% noncondensing

Interfaces:

USB

RS-232

USB 2.0, 480 Mbps

3-wire RS-232

 