QE65000 Spectrometer
QE65000 Spectrometer
QE65000 Spectrometer
QE65000
Description
The Ocean Optics QE65000 Spectrometer includes the linear CCD-array optical bench, plus all the circuits necessary for spectrometer operation. The result is a compact, flexible system, with no moving parts, that's easily integrated as an OEM component
The QE65000 spectrometer is a unique combination of technologies providing users with both an unusually high spectral response and high optical resolution in a small footprint. The electronics have been designed for considerable flexibility in connecting to various modules as well as external interfaces. The QE65000 interfaces to PCs, PLCs and other embedded controllers through USB 2.0 or RS-232 communications. The information included in this data sheet provides detailed instructions on the connection and operation of the QE65000.
The detector used in the QE65000 spectrometer is a scientific-grade, back-thinned, TE Cooled (TEC), 1044x64 element CCD array from Hamamatsu (product number S7031-1006). For complete details on this detector, visit www.Hamamatsu.com.
The QE65000 operates from power provided through the USB (without TEC operation), or from a separate + 5VDC power supply. The QE65000 is a microcontroller-controlled spectrometer, thus all operating parameters are implemented through software interfacing to the unit.
Features
Hamamatsu S7031-1006 Detector:
Spectrometer Design:
Electrical Performance:
Embedded microcontroller allows programmatic control of all operating parameters and standalone operation:
TE Cooler
Onboard Pulse Generator:
Onboard GPIO:
EEPROM storage for:
Plug-n-Play Interface for PC applications
30-pin connector for interfacing to external products
Specifications:
Specifications |
Criteria |
Physical Specifications: Physical Dimensions (LxWxH) Spectrometer Weight Power Supply Weight |
182 mm (7.17 in.) x 110 mm (4.33 in.) x 47 mm (1.85 in.) 1.15 kg (2.6 lbs.) 0.45 kg (1 lb.) |
Power: Power requirement Supply voltage Power-up time GPIO Output Voltage |
500 mA at +5 VDC (no TE Cooler) 3.5A at +5V (with TE Cooler) 4.5 – 5.5 V ~2s depending on code size 3.3 V |
Absolute Maximum Ratings: VCC Voltage on GPIO pin |
+ 5.5 VDC 3.6 V |
Spectrometer: Design Focal length (input) Input Fiber Connector Gratings Entrance Slit Detector Pixels Pixel size Active area Spectral range Quantum efficiency Well Depth Dark Current Filters |
Symmetric crossed Czerny-Turner F/4 101mm SMA 905 to single-strand optical fiber (0.22 NA) 14 different gratings (H1 – H14); HC1 grating 5, 10, 25, 50, 100, or 200 μm slits. (Slits are optional. In the absence of a slit, the fiber acts as the entrance slit.) Hamamatsu S7031-1006 1024 x 58 (total pixels 1044 x 64) 24.6μm square 24.576 x 1.392mm 200 – 1100nm with window (see etalon section). VUV option available (150nm). VUV option includes purge port & window removal (no TE Cooling) 90% peak (see figure following this table) Individual pixel: 300K electrons; Column summing well: ~1.5M electrons 4000 e/pixel/sec (typ) @ 25° C; 200 e/pixel/sec (typ) @ 0° C 6 OF Series available |
Spectroscopic: Integration Time Dynamic Range Signal-to-Noise Dark Noise (single dark spectrum) Readout Noise Linearity |
8ms – 15 minutes 40000:1 single integration period, 7.5 x 109 over integration range 1000:1 single acquisition 3 counts RMS 1.5 counts RMS (40 electrons) ~1% max. nonlinearity (uncorrected); <0.5% nonlinearity (corrected) |
Specifications |
Criteria |
Environmental Conditions: Temperature TEC Control Temperature Humidity |
0 to 50°C 45 to 30°C below ambient temperature 0% – 90% noncondensing |
Interfaces: USB RS-232 |
USB 2.0, 480 Mbps 3-wire RS-232 |