Categories

High resolution spectrometer

High resolution spectrometer

High resolution spectrometer HR4000

Description

The Ocean Optics HR4000 Spectrometer includes the linear CCD-array optical bench, plus all the circuits necessary for spectrometer operation. The result is a compact, flexible system, with no moving parts, that's easily integrated as an OEM component.

 

 

 The HR4000 spectrometer is a unique combination of technologies providing users with both an unusually high spectral response and high optical resolution in a single package. The electronics have been designed for considerable flexibility in connecting to various modules as well as external interfaces. The HR4000 interfaces to PC’s, PLC’s and other embedded controllers through USB 2.0 or RS-232 communications. The information included in this guide provides detailed instructions on the connection and operation of the HR4000.

The detector used in the HR4000 spectrometer is a high-sensitivity 3648-element CCD array from Toshiba, product number TCD1304AP. Ocean Optics applies a coating to all TCD1304AP detectors, so the optical sensitivity could vary from that specified in the Toshiba datasheet).

 

The HR4000 operates off of a single +5VDC supply and either a USB or RS-232 interface. The HR4000 is a microcontroller-controlled spectrometer, thus all operating parameters are implemented through software interfacing to the unit.

 

Features:

 

q TCD1304AP Detector

 High sensitivity detector

 Readout Rate: 1MHz

 Shutter mode

q Optics

 An optical resolution of ~0.03nm (FWHM)

 A wide variety of optics available

 14 gratings

 6 slit widths

 1 order sorting filter

q Electrical Performance

 14 bit, 5MHz A/D Converter

 Integration times from 3.8ms to 10s

q 4 triggering modes

q Embedded microcontroller allows programmatic control of all operating parameters & Standalone operation

 USB 2.0 480Mbps (High Speed) & 12Mbps (Full speed)

 RS232 115Kbaud

 Multiple Communication Standards for digital accessories (SPI, I2C)

q Onboard Pulse Generator

 2 programmable strobe signals for triggering other devices

 Software control of nearly all pulse parameters

q Onboard GPIO

 10 user programmable digital I/O

q Onboard Analog Interface

 Analog Output: 9bit, 0-5V

q EEPROM storage for

 Wavelength Calibration Coefficients

 Linearity Correction Coefficients

 Absolute Irradiance Calibration (optional)

q Plug-n-Play Interface for PC applications

q 30-pin connector for interfacing to external products

q CE Certification

 

Specifications

Specifications

Criteria

Absolute Maximum Ratings:

VCC

Voltage on any pin

+ 5.5 VDC

Vcc

Physical Specifications:

Physical Dimensions

Weight

148.6 mm x 104.8 mm x 45.1 mm

570 g

Power:

Power requirement (master)

Supply voltage

Power-up time

500 mA at +5 VDC

4.5 – 5.5 V

~5s depending on code size

Spectrometer:

Design

Focal length (input)

Focal length (output)

Input Fiber Connector

Gratings

Entrance Slit

Detector

Filters

Asymmetric crossed Czerny-Turner

F/4 101mm

68 mm (75, 83 and 90 mm focal lengths also available)

SMA 905 to single-strand optical fiber (0.22 NA)

14 different gratings

5, 10, 25, 50, 100, or 200 μm slits. (Slits are optional. In the absence of a slit, the fiber acts as the entrance slit.)

Toshiba TCD1304AP

2nd & 3rd order rejection, long pass (optional)

Spectroscopic:

Integration Time

Dynamic Range

Signal-to-Noise

Readout Noise (single dark spectrum)

Resolution

Spectrometer Channels

3.8 ms – 10sec

2500 single integration period, 2.5 x 109 over integration range

300:1 single acquisition

6 counts RMS, 20 counts peak-to-peak

0.03 – 1.0 nm varies by configuration

One

Environmental Conditions:

Temperature

Humidity

-30 to +70 C Storage & -10 to +50 C Operation

0% - 90% non-condensing

Interfaces:

USB

RS-232

I2C

USB 2.0, 480 Mbps

2-wire RS-232

Inter-Integrated Circuit 2-Wire serial BUS

 